18
Aesculap
®
Lektrafuse HF generator GN200
7.6
User maintenance
Ź
In case of error, follow the instructions in the display and note the error
code.
Ź
If the error continues after following the instructions in the display,
contact your national B. Braun/Aesculap agency, see Technical Service.
7.7
Fuse replacement
Specified fuses: 2 units Aesculap art. no. TA021404:
G fuse link, time lag (T) 8 A, breaking capacity H (1 500 A)
Ź
Use a small screwdriver to release the clip on the fuse holder 11 and
tilt up cover.
Ź
Use a small screwdriver to release the fuse holder 11 and pull it out.
Ź
Replace both fuses.
Ź
Reinsert the fuse holder 11.
Ź
Close the cover.
Fig. 14
Note
If the fuses burn out frequently, the device is faulty and should be
repaired, see Technical Service.
8.
Technical Service
Ź
For service and repairs, please contact your national B. Braun/Aesculap
agency.
Modifications carried out on medical technical equipment may result in
loss of guarantee/warranty rights and forfeiture of applicable licenses.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:
+49 (7461) 95 -1601
Fax:
+49 (7461) 14 -939
E-Mail:
ats@aesculap.de
Or in the US:
Aesculap Inc.
Attn. Aesculap Technical Services
615 Lambert Pointe Drive
Hazelwood
MO, 63042 USA
Aesculap Repair Hotline
Phone:
+1 (800) 214 -3392
Fax:
+1 (314) 895 -4420
Other service addresses can be obtained from the address indicated above.
9.
Accessories/Spare parts
Note
Other accessories, applied parts and replacement parts are described in
Aesculap brochure C-304-81 and C-902-02
10. Technical data
Classification acc. to Directive 93/42/EEC
DANGER
Risk of death by electric shock!
Ź
Unplug the device before changing the fuses!
DANGER
Danger to life of patients and users if the product
malfunctions and/or protective measures fail or are
not used!
Ź
Do not perform any servicing or maintenance
work under any circumstances while the product
is being used on a patient.
Ź
Do not modify the product.
Art. no.
Designation
TA021404
Fuse
–
Caiman 5 instrument family
–
Caiman 12 instrument family
GN201
Foot switch
Art. no.
Designation
Class
GN200
High frequency surgical device
IIb