Lektrafuse HF generator GN200
2.5
Acoustic warning signal
2.6
Monitoring functions
Self-test
As soon as it is switched on, the unit performs a self-test of the control
elements, the acoustic warning signal, the microprocessor and the hard-
ware function. During this phase, the message "Performing Self-Test"
appears in the display 5.
Fig. 1
Continuous test cycle during operation
During operation, safety-relevant functions and signals are monitored
through a continuous test cycle. As soon as a critical error is detected, the
HF generator terminates the HF activation. In display 5, the error number
of the corresponding error is displayed and an acoustic warning signal is
emitted, see Troubleshooting list.
Status
Acoustic warning signal
Notes
System failure
3 repetitions of: 1 long beep, 3 KHz, 1 beep, 2 KHz Following each system failure, if identified (F001,...)
REGRASP error
3 repetitions of: 1 beep, 2 KHz,
1 beep, 3 KHz
Following each regrasp error, if identified
System error
3 beeps, 3 KHz
Following each failure report, if identified (E001,...)
System warning
3 beeps, 3 KHz
Following each warning, if identified
Start HF discharge
2 beeps, 480 Hz
When starting the HF discharge
HF discharge active
1 beep, 480 Hz
Ŷ
In Standard mode: Repetition with 1.8 Hz
Ŷ
In Plus mode: Repetition with 2.8 Hz
Continually during HF discharge
End of HF discharge
3 beeps, 530 Hz
When a sealing process has been successfully completed (no
failure reports or errors)
Power up self-test
1 beep, 2 KHz
During the self-test, in order to check that the signal tone is
functioning