Easy At Home EBP-017 User Manual - page 11
Power Supply
Battery powered mode:
2×AAA batteries (3V)
Display Mode
Measurement Mode
Oscillographic testing mode
Measurement Range
Accuracy
Normal working condition
Storage & transportation
condition
Measurement perimeter
of the wrist
About 13.5cm-21.5cm( )
Weight
External dimensions
Attachment
2×AAA batteries,one storage,user manual
Mode of operation
Continuous operation
Degree of protection
Software version
Device classification
V01
Internally Powered ME Equipment
WARNING: No modification of this equipment is allowed.
Contact Information
Complied Standards List
For more information about our products, please visit ,
or call Customer 1-866-822-6999 M-F 9 a.m.-5 p.m.CST, usual
problems and customer download, will serve you anytime.
Pressure: 0kPa-40kPa(0mmHg-300mmHg)
pulse value:(40-199)beat/minute
Approx.112g(3.95oz)(Excluding batteries)
Type B applied part
IP Classification
IP×0
19
18
CONTACT INFORMATION
SPECIFICATIONS
Risk management
Labeling
User manual
General Requirements
for Safety
Electromagnetic
compatibility
Non-invasive
Sphygmomanometers
General
Requirements
ISO 14971:2007
EN 980:2008
EN 1041:2008
IEC 60601-1:2005
IEC 60601-1-2:2007
AASI/AAMI SP10:2002/A1:2003/A2:2006/(R)2008
Digital LCD V.A 31.5mm×44mm(1.25’’×1.73’’)
Pressure:
5℃-40℃(41℉-104℉)within±0.4kPa(3mmHg)
pulse value:±5%
Temperature:5℃ to 40℃(41℉ to 104℉)
Relative humidity: ≤85%RH
Atmospheric pressure: 86kPa to 106kPa
Temperature:-20℃-60℃
Relative Humidity: 10%RH-93%RH
Atmospheric Pressure: 50kPa-106 kPa
(-4℉ - 140℉)
Approx. 68mm×75mm×31mm(2.68”×2.95”×1.22”)
(Excluding the cuff)