Easy At Home EBP-017 User Manual - page 3
3
2
INTRODUCTION
INTRODUCTION
LCD Display Signal
Systolic
blood pressure
High pressure result
Diastolic
blood pressure
Low pressure result
BPM
Pulse/minute; heartbeats/minute
Memory
mmHg
Measurement unit the blood pressure
(1mmHg=0.133kPa)
kPa
Measurement unit of the blood pressure
(1kPa=7.5mmHg)
Year/Month/Day,Hour/Minute
Current time
Movement error
symbol
Shocking will result in inaccurate
CUFF air is exhausting of deflating
Deflating
Low battery
Batteries are low and need to be replaced
Grade
The grade of the blood pressure
SYMBOL DESCRIPTION EXPLANATION
Arrhythmia
Irregular heartbeat
Heartbeat
Heartbeat detection during the
measurement
The displayed measurement values
is from the memory.
This device is intended for adult use only.
This device is intended for non-invasive measuring and monitoring of arterial blood pressure. It is
not intended for use on extremities other than the wrist or for functions other than obtaining a blood
pressure measurement.
Do not confuse self-monitoring with self-diagnosis. This unit allows you to monitor your blood
pressure.Do not begin or end medical treatment without asking a physician for treatment advice.
If you are taking medication,consult your physician to determine the most appropriate time to
measure your blood pressure. Never change a prescribed medication without consulting your
physician.
If the cuff pressure exceeds 40 kPa (300 mmHg),the unit will automatically deflate. Should the cuff
not deflate when pressures exceeds 40 kPa (300 mmHg), detach the cuff from the wrist and press
the START/STOP button to stop inflation.
To avoid measurement errors, carefully read this manual before using the product.
The equipment is not AP/APG equipment and not suitable for use in the presence of a flammable
anesthetic mixture with air of with oxygen or nitrous oxide.
The operator shall not touch output of batteries and the patient simultaneously.
The user must check that the equipment functions safely and see that it is in proper working
condition before being used.
CAUTION
This device is contraindicated for any female subject who may be suspected of, or is pregnant.
Besides providing inaccurate readings, the effects of this device on the fetus are unknown.
Please use ACCESSORIES and detachable partes specified/ authorised by MANUFACTURE.
Otherwise, it may cause damage to the unit or danger to the user/patients.
Manufacturer will make available on request circuit diagrams, component parts list etc.
This unit is not suitable for continuous monitoring during medical emergencies or operations.
Otherwise, the patient’s wrist and fingers will become anaesthetic, swollen and even purple due to a
lack of blood.
Please use the device under the environment which was provided in the user manual. Otherwise,
the performance and lifetime of the device will be impacted and reduced.
During use, the patient will be in contact with the cuff.The materials of the cuff have been tested and
found to comply with requirements of ISO 10993-5:2009 and ISO 10993-10:2010. It will not cause any
potential alergic reaction or contact injury.
The device doesn’t need to be calibrated within the two years of reliable service.
Please dispose of ACCESSORIES, detachable parts, and the ME EQUIPMENT according to the
local guidelines.
When the device was used to measure patients who have common arrhythmias such as atrial or
ventricular premature beats or atrial fibrillation, the best result may occur with deviation. Please consult
your physician about the result.
The device has been evaluated clinically using manual cuff/stethoscope auscultation as the reference.
Blood pressure measurements determined with this device are equivalent to those obtained by a
trained observer using the cuff/stethoscope auscultatory method, within the limits prescribed by the
American National Standard, Manual, electronic, or automated sphygmomanometers.”
Do not use the monitor under the conditions of strong electromagnetic field (e.g. medical RF
equipment) that radiates interference signal or electrical fast transient / burst signal.