TROUBLESHOOTING
This section includes a list of error messages and frequently
asked questions for problems you may encounter with your blood
pressure monitor. If the products not operating as you think it
should, check here before arranging for servicing.
PROBLEM SYMPTOM
CHECK THIS
REMEDY
No power
Low
batteries
Error
massage
Display is dim or
will not light up.
Batteries are exhausted.
Replace with new batteries
Insert the batteries correctly
Replace with new batteries
Batteries are inserted
incorrectly.
Show on the display
Batteries are low.
E 1 shows
The cuff is not
secure.
Refasten the cuff and then
measure again.
E 2 shows
The cuff is very tight
Refasten the cuff and then
measure again.
E 3 shows
The pressure of the
cuff is excess.
Relax for a moment
and then measure again.
E 10 or E11
shows
The monitor detected
motion while measuring.
Movement can affect the
measurement.Relax for a
moment and then measure
again.
E20 shows
EExx,shows on
the display.
A calibration error
occurred.
Retake the measurement.If
the problem persists,contact
the retailer or our customer
service department for
further assistance.
Refer to the warranty for
contact information and
return instructions.
AC adaptor is inserted
incorrectly.
Insert the AC adaptor
tightly
The measurement
process does not detect
the pulse signal.
Loosen the clothing on the
arm and then measure
again
E21 shows
The treatment of the
measurement failed.
Relax for a moment and
then measure again.
SPECIFICATIONS
Power supply
Battery powered mode:
6VDC 4*AA batteries
AC adaptor powered mode:
(INPUT: 100-240VAC 50/60Hz 0.3A Max
OUTPUT: 6V 1000mA)(Not Included)
Display mode
Digital LCD V.A.60mm*80mm(2.36’’*3.15’’)
Measurement mode
Oscillographic testing mode
Measurement range
Accuracy
Normal working condition
Storage & transportation
condition
Measurement perimeter
of the upper arm
About 22cm~32cm( )
Net Weight
Approx.388g(13.69oz)(Excluding the batteries)
External dimensions
Attachment
Approx.102mm*143mm*73mm
(4.02”*5.63”*2.87”)
4*AA batteries, one storage bag, user manual
Mode of operation
Continuous operation
Degree of protection
Software Version
Type B applied part
Protection against
ingress of water
V01
Pressure: 0mmHg~300mmHg(0kPa-40kPa)
pulse value:(40-199)times/minute
IPX0
Pressure:
5℃-40℃(41℉-104℉)within±3mmHg(0.4kPa)
pulse value:±5%
Temperature:5℃ to 40℃(41℉ to 104℉)
Relative humidity: ≤85%RH
Atmospheric pressure: 86kPa to 106kPa
Temperature:-20℃-60℃
(-
4℉ to 140℉)
Relative Humidity: 10%RH-93%RH
Atmospheric Pressure: 50kPa-106 kPa