EDAN iM80 Servise Manual - page 18
iM80 Patient Monitor Service Manual Test and Maintenance
- 12 -
4.2.2 SpO
2
Performance Test
This test checks the performance of the SpO
2
measurement.
Tools required: SpO
2
simulator.
Procedure:
1. Connect the monitor and the SpO
2
simulator with a SpO
2
cable.
2. Switch on the monitor and the simulator.
3. Set the simulator to the following configuration:
- SpO
2
= 70%.
4. Check the displayed SpO
2
value against the simulator configuration. The value should be
70% ±2%.
4.2.3 NIBP Performance Test
This test checks the performance of the NIBP measurement.
Tools required:
• NIBP simulator;
• T-fitting;
• Extension tube;
• Artificial limb.
Procedure:
1. Connect the NIBP simulator to the monitor.
2. Switch on the monitor and the simulator. Calibrate the simulator before using it.
3. Set the patient type on the monitor to adult; set the simulator to the following configuration:
- Patient type: adult;
- Systolic pressure=255mmHg;
- Diastolic pressure=195mmHg;
- Mean pressure=215mmHg.
And then start a NIBP measurement.
4. Check the displayed values against the simulator configuration. The differences should be