iM80 Patient Monitor Service Manual Configuration
- 23 -
MFM-CMS central monitoring system by EDAN. To set the device No., the user should:
1
Enter User Maintain;
2
Select Network Maintain > Bed No.;
3
Select a device No. from 1 to 254 and press the knob to confirm it.
CAUTION
Make sure the device No. of the monitors in the same system do not overlap.
5.5 Changing Network Bed No.
This network bed No. is the physical device No. of the monitor in the network and used for
setting the IP address of the device. It needs to be set if the device connects to the CMS network.