iM80 Patient Monitor Service Manual Disassembling the Monitor
- 51 -
Assemble the power interface board in the reverse order.
8.6.6 Replacing the Interface Board (big)
To disassemble the
Interface Board (big)
:
Separate main bracket
assembly as described in
section 8.3
Disconnect
wires
connecting interface
board (big)
Unscrew the screw
securing back board
Unscrew the screws and
nuts securing power
interface board (big)
Separate main bracket
assembly as described in
section 8.3
Disconnect wires connecting power
interface board and unscrew the
screws
Unscrew the screws and nuts
securing power interface board