EDAN iM80 Servise Manual - page 20
iM80 Patient Monitor Service Manual Test and Maintenance
- 14 -
• Cylinder;
• Manometer.
Procedure:
1. Access Menu > Maintenance > User Maintain by inputting the password ABC.
2. Connect the equipment as shown below:
Manometer
NIBP module
/ Patient
monitor
Cylinder
T-fitting
NIBP
extension
tube or hose
Tube
Tube
Figure 4-2 Diagram for NIBP Calibration
3. Select NIBP Maintain > Calibrate.
4. Apply fixed static pressure on the monitor with the help of the manometer. Compare the
displayed values on the monitor with the manometer values.
5. A difference within the range of ±3mmHg is reasonable.
4.2.6 TEMP Performance Test
This test checks the performance of the TEMP measurement.
Tools required: resistance box.
Procedure:
1. Switch on the monitor and the resistance box.
2. Set the probe type on the monitor to YSI-10K, and respectively connect the probes to
channel T1 and T2 connectors. And then connect the probes with the resistance box.
3. Set the resistance value to (6017
Ω) 37℃ in the resistance box.
4. A difference within the range of ±0.1℃ is reasonable.