iM80 Patient Monitor Service Manual Disassembling the Monitor
- 40 -
3
Put the monitor upward and slightly pull the front and back housings, and disconnect the
wires to LCD and key board to separate the front housing and back housing assemblies,
shown in circle in the figure below;
4
Place the back assembly on a flat surface, and unscrew the screws securing the module
bracket assembly and disconnect the wires to separate the module bracket assembly;
5
Unscrew the screws securing the main frame on the bottom and inside the unit, and the main
frame are separated from the back assembly.
Screws:
M3X12