Comparing the Control Solution Test Results
The test result of each control solution should be within the
range printed on the label of the test strip vial. Repeat the
control solution test if the test result falls outside of the range.
Out of range results may occur due to the following factors:
If results continue to fall outside the range printed on the
test strip vial, the CareSens N Test Strip and Meter may not be
working properly. Do not use your system and contact your
authorised i-SENS sales representative.
Situations
•Whenthecontrolsolutionbottlewas
not shaken well,
•Whenthemeter,teststrip,orthe
control solution were exposed to high
or low temperatures,
•Whenthefirstdropofthecontrol
solution was not discarded or the tip of
the bottle was not wiped clean,
•Whenthemeterisnotfunctioning
properly.
•Whenthecontrolsolutionispastthe
expiration date printed on the bottle,
•Whenthecontrolsolutionispastits
discard date (the date the bottle was
opened plus three months),
•Whenthecontrolsolutionis
contaminated.
Actions
Repeat the
control solution
test by referring
to the “Notes”
on page 20.
Discard the used
control solution
and repeat the
test using a new
bottle of control
solution.
Step 3
The display segments will rotate
clockwise on the meter display and a test
result will appear after the meter counts
down from 5 to 1. After your control
solution result appears on the display,
press
for 3 seconds till the "check"
appears on the display. When the "check"
is displayed, the result is not stored in the
meter’s memory and is not included in
the averages.
Step 4
Compare the result displayed on the meter
to the range printed on the test strip vial.
The result should fall within the range.
Used strips should be discarded safely in
appropriate containers.
Caution:
The range printed on the test strip vial is for the
CareSens Control Solution only. It does not have any
connection to your blood glucose level.
Note:
The CareSens Control Solution can be purchased
separately. Please contact your authorised i-SENS sales
representative.
Control Solution Range
Control A: 5.6~8.4 mmol/L
Control B:10.2~15.3 mmol/L