I-SENS CareSens N Pop Manual - page 13
26 www.i-sens.com
www.i-sens.com 27
Step 7
Insert a test strip with the contact bars facing upwards into the
meter’s test strip port. Push the strip in gently until the meter
beeps. Be careful not to bend the test strip. The symbol will
appear on the screen.
Step 8
If you want to attach a post-meal flag
( ) to a test result, press and hold
for 3 seconds after inserting
the test strip. The post-meal flag ( )
and the symbol will appear on the screen.
The test result will also be displayed with the post-meal flag
( ). If you do not want to save the result as a post-meal
test, move on to step 9 after step 7.
Preparing the Meter and Test Strip
Flagging Post-meal Test Results
The CareSens N POP meter allows you to flag a result of a
post-meal test with symbol. The post-meal test flag
( ) can be attached just before applying the blood sample.
Once you attach the post-meal flag ( ) to the test results, it
cannot be deleted.
Step 5
Select a desired depth of one-to-seven
(1-7) on lancing device's adjustable tip.
Choose a depth by rotating the top
portion of the adjustable tip until the
setting number matches the arrow.
Step 6
To cock the lancing device, hold the
body in one hand and pull the sliding
barrel with the other hand. The device
is loaded when you feel a click and the
load confirmation window turns red.
Note:
The skin depth to get blood samples will vary for
various people at different sample sites. The lancing device's
adjustable tip allows the best depth of skin penetration to get
an adequate sample size.
Note:
1 = least penetration of lancet into the skin
7 = most penetration of lancet into the skin
3
4
5
6
7
1
2
3
4
5
6
7