I-SENS CareSens N Voice User Manual - page 10
16 www.i-sens.com
www.i-sens.com 17
Step 9
Setting or Turning off the Sound
A number showing the sound volume will blink and the meter
will say "Set volume." Press ▲ or ▼ button to adjust the volume
from 1 (lowest) to 3 (highest) or to turn the sound off or the beep
sound on.
When you select the beep On the meter will beep instead of
announcing messages. To turn off the sound, press the S button
when OFF blinks on the display. Then the symbol will appear
on the display and the meter will be muted. Press the S button
to save the setting and move to the hypoglycemia (low blood
glucose) indicator setting mode.
Note:
Only when the sound is set to OFF, the symbol
appears on the display.
Setting the Hypoglycemia Indicator
You can set the meter to let you know when your test result
indicates a possible low blood glucose (hypoglycemia). You can
also select the hypoglycemia level from 60 to 20 mg/dL.
Step 10
Turning on the Hypoglycemia Indicator
When you enter the hypoglycemia indicator setting mode, the
symbol and HYPo will blink. If the talking function is on, the
meter will say "Set hypoglycemia
indicator." On pressing ▲ or ▼, the
screen will display On or OFF. Press
S button to set your hypoglycemia
level if you choose On. The screen
will return to step 2, language
selection mode, if you choose OFF.
Caution:
Ask your healthcare professional to help you
decide what your hypoglycemia level is before setting your
hypoglycemic level.
Step 11
Setting Your Hypoglycemic Level
The hypoglycemic level can be set
from 60 to 20 mg/dL. To set your
hypoglycemic level, press ▲ or ▼
button until the level you want to
appear. Press the S button to confirm.
Then the screen will return to step 2,
language selection mode (See page 13).
Step 12
Exiting from Set Mode
After completing meter setting, press and hold the S button
for 3 seconds to confirm all your selections and exit from SET
mode. If the talking function is on, the meter will announce the
set date and time for your confirmation.