ACCESSORIES
TECHNICAL INFORMATION
Type: TA011
Dimensions: 24 - 29 mm (W) * 35 - 95 mm (H) * 215 mm (L).
Weight: 140 g (without batteries), 230 g (4 x AA batteries).
Power supply: 4 x AA non-rechargeable batteries, 1.5V alkaline LR3.
Measurement range: 7-50 mmHg
Accuracy: ±1.2 mmHg (≤20 mmHg) and ±2.2 mmHg (>20 mmHg).
Repeatability (coefficient of variation): <8 %.
Accuracy of display: 1 mmHg.
Display unit: Millimeter mercury (mmHg).
The serial number is on the inside of the battery compartment cover.
There are no electrical connections from the tonometer to the patient.
The device has BF-type electric shock protection.
Operation environment:
Temperature: +10 °C to +35 °C
Relative humidity: 30 % to 90 %
Atmospheric pressure: 800 hPa-1,060 hPa
Storage environment:
Temperature: -10 °C to +55 °C
Relative humidity: 10 % to 95 %
Atmospheric pressure: 700 hPa-1,060 hPa
Transport environment:
Temperature: -40 °C to +70 °C
Relative humidity: 10 % to 95 %
Atmospheric pressure: 500 hPa-1,060 hPa
Mode of operation: continuous
SKU
PRODUCTDESCRIPTION
WEIGHT DIMENSIONS
104
Box of 100 probes
89 g
53 x 109 x 36 mm
7217
Probe base collar, Icare ic100
4 g
18 x 18 mm
540
Probe base
1 g
7 x 38 mm
559
Wrist strap with lock
4 g
10 x 10 x 270 mm
525
Aluminium case, Icare ic100
800 g
240 x 280 x 72 mm
7169
Battery cover & screw, Icare ic100
6 g
110 x 25 x 12 mm
623
IOP pad, Icare ic100
40 g
50 x 56 x 16 mm
543
Probe base cleaning container
3 g
20 x 56 mm
565
Silicon grip - white, Icare ic100
26 g
45 x 35 x 113 mm
548
Screw driver, Icare
15 g
16 x 90 mm
577
USB memory stick
44 g
98 x 11 x 93 mm
544
Probe base plug
1 g
19 x 11 mm
PERFORMANCE DATA
The performance data is obtained from a clinical study, performed according to American National Standard ANSI Z80.10-2003 and International Standard ISO
8612.2 for tonometers. The study was performed ain the Department of Ophthalmology, Helsinki University Central Hospital. In the study, 158 patients were
measured. The mean paired difference and standard deviation (Goldmann-Icare) were -0.4 mmHg and 3.4 mmHg. A scattergram and Bland-Altman plot of the
results is shown below.