MEASUREMENT
NOTE
If you do not use the tonometer, it will switch off automatically after 3 minutes.
NOTE
No anesthetic is required when performing measurements. *Since local anesthetic may lower the tonometer
reading, we recommend that you refrain from using an anesthetic when performing measurements.
STEP 1.
Ask the patient to relax and look straight ahead at a specific point. Bring the tonometer near the patient’s eye.
Correct alignment of the tonometer and green probe base light indication.
Incorrect alignment of the tonometer and red probe base light indication.
Correct head and eye position.
Incorrect head and eye position.
STEP 2.
The device should be in a horizontal position. Keep the probe horizontal and pointing perpendicularly to the center of the cornea. The distance from the tip
of the probe to the patient’s cornea (see picture) should be 4-8 mm (5/32 - 5/16 inch).
If necessary, adjust the distance
by turning the forehead support
adjusting wheel.
If probe base light indication is set OFF
green arrows will indicate the correct
alignment of the tonometer.
If probe base light indication is set OFF
red arrows will indicate the incorrect
alignment of the tonometer.