Icare ic100 Instruction Manual - page 14
Icare ic100 tonometer manual
ENGLISH
www.icaretonometer.com
14
TA011 is class B equipment and needs special precautions regarding EMC and needs to be installed and put into service according to EMC information provided in
user and maintenance manual.
GUIDANCEANDMANUFACTURER’SDECLARATION–ELECTROMAGNETICEMISSIONS
Icare ic100 (TA011) is intended for use in the electromagnetic environment specified below.
The user of the Icare ic100 (TA011) should assure that it is used in such an environment.
RF emissions CISPR 11
Group 1
Icare ic100 (TA011) is battery operated and use RF energy only for its internal function. Therefore,
its RF emissions are low and are not likely to cause any interference in nearby equipment.
RF emissions CISPR 11
Class B
Icare ic100 (TA011) is suitable for use in all establishments, including domestic establishments and
those directly connected to public low-voltage power supply network that supplies buildings used
for domestic purposes
Harmonic emissions
IEC 61000-3-2
NOT APPLICABLE
Voltage fluctuations flickering
emissions
IEC 61000-3-3
NOT APPLICABLE
GUIDANCEANDMANUFACTURER’SDECLARATION–ELECTROMAGNETICIMMUNITY
Icare ic100 (TA011) is intended for use in the electromagnetic enviroment specified below.
The customers or users of Icare ic100 (TA011)) should assurre that it is used in such enviroment
Immunity test
IEC 60601 Test level
Compliance level
Electromagnetic environment-Guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
± 6 kV contact
±8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile. If floors are covered with
synthetic material, the relative humidity should be at least 30 %
Electrical fast Transients/burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
NOT APPLICABLE
Surge
IEC 61000-4-5
±1 kV for line(s) to line(s)
±2 kV for line(s) to earth
NOT APPLICABLE
Voltage dips, short interruption
and voltage variations on power
supply lines
IEC 61000-4-11
<5 % UT (>95 % dip in UT)
for 0,5 cycle
40 % UT (60 % dip in UT)
for 5 cycles
70 % UT (30 % dip in UT)
for 25 cycles
<5 % UT (>95 % dip in UT)
for 5 s
NOT APPLICABLE
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels characteristic of a
typical location in a typical commercial or hospital environment.