MESSAGE
DESCRIPTION
ACTIONS
Battery charge is low.
Prepare to replace the batteries.
The batteries are empty.
Turn the tonometer OFF by pressing Select button.
Replace the batteries.
The probe did not move.
Change the probe. The probe was twisted or otherwise inserted incorrectly.
To clear error messages, press the Measure button, after which the
measurement can be repeated.
The probe did not move properly for several times during
the measurement sequence.
Remove and clean the probe base or replace it with new one as instructed in
Replacing/cleaning the probe base. To clear error messages, press the Mea-
sure button, after which the measurement can be repeated.
The probe did not touch the eye.
Adjust correct measurement distance 4-8 mm. The measurement was taken
from too far away.
Too short measurement distance between the probe and
the cornea.
Adjust correct measurement distance 4-8 mm. The measurement was taken
from too close. To clear error messages, press the Measure button, after
which the measurement can be repeated.
The probe did not move properly. The probe did not make
clean contact with the cornea, because the probe hit an
eyelid or eyelashes.
Ensure that the eye is open, measure again. To clear error messages, press the
Measure button, after which the measurement can be repeated.
Internal error detected.
Turn the tonometer OFF by pressing Select button. Contact the seller to
arrange sending the device for service.