NBN 9320 User Manual

Manual is about: Load Cell Hand Held Digital Indicator, TEDS Plug & Play Ready, IEEE 1451.4 Compliant

Summary of 9320

  • Page 1

    9320 load cell hand held digital indicator teds plug & play ready ieee 1451.4 compliant user manual 13-69 rev 2.4 interface, inc. • 7401 e. Butherus dr. • scottsdale, az 85260 • 800-947-5598 www.Interfaceforce.Com 9320tedsman2r2_4.Doc nbn elektronik handelsgesellschaft m. B. H. Riesstraße 146, 8010 ...

  • Page 2: Contents

    Interface 9320 teds ready indicator manual issue 2.4 2 contents what is teds? ................................................................................................ 3 basic concept ................................................................................................ 3 how it wor...

  • Page 3: What Is Teds?

    Interface 9320 teds ready indicator manual issue 2.4 3 what is teds? "plug and play sensor hardware and software make configuring a smart teds sensor as easy as plugging a mouse into a pc. The technology has greatly improved efficiency and productivity by completely eliminating manual sensor configu...

  • Page 4: Introduction

    Interface 9320 teds ready indicator manual issue 2.4 4 advantages plug and play sensors are revolutionizing measurement and automation. With transducer electronic data sheets (teds), your data acquisition system can detect and automatically configure sensors. This technology provides: • reduced conf...

  • Page 5

    Interface 9320 teds ready indicator manual issue 2.4 5 electrical connection information sensor connections the standard sensor connection is a 5 pin 723 series binder connector. The sensor connector is also used as a charging socket, when the re-chargeable alkaline battery option has been ordered, ...

  • Page 6

    Interface 9320 teds ready indicator manual issue 2.4 6 there are six push buttons on the front panel of the 9320, which are available for use in normal operation. Each of these is described below: front panel button function of button in normal operation mode to switch the 9320 on or off press and h...

  • Page 7

    Interface 9320 teds ready indicator manual issue 2.4 7 menu structure the 9320 has two menus, details of which are outlined below: a configuration menu, which enables the user to tailor the operation to meet a specific application requirement. The values selected in the configuration menu are comple...

  • Page 8: Set Zero

    Interface 9320 teds ready indicator manual issue 2.4 8 sens 5.0 set res live? 0000.000 use sc? Apply lo disp lo 0000000 apply hi disp hi 0000000 done apply lo disp lo 0000000 disp hi 0000000 done calibrat table? Input lo 0000000 disp lo 0000000 input hi 0000000 disp hi 0000000 done cal val? Set 9ain...

  • Page 9: Set

    Interface 9320 teds ready indicator manual issue 2.4 9 this allows the setting of a visual overload. The value entered is the display value at which the 9320 displays ouerload . Values between -9999999 and +9999999 can be entered, using the and arrows to select a digit and the and arrows to incremen...

  • Page 10: Sens 5.0

    Interface 9320 teds ready indicator manual issue 2.4 10 calibration menu to enter the calibration menu, press and hold and buttons for 5 seconds parameter set-up information sens 5.0 press to skip to next menu item press to change sensor input sensitivity this allows the calibration engineer to chan...

  • Page 11: Teds

    Interface 9320 teds ready indicator manual issue 2.4 11 teds enabling teds disables calibrate menu press to skip this parameter and exit menu press to enable or disable teds. If you have chosen to enter the teds calibration, enabled? Appears. If you have chosen to enter the teds you will be prompted...

  • Page 12

    Interface 9320 teds ready indicator manual issue 2.4 12 hold button the hold button allows the user to freeze the display when it is pressed. When pressed again the display returns to it’s normal operating mode. When in hold mode the display will flash and the hold annunciator will be lit, to ensure...

  • Page 13

    Interface 9320 teds ready indicator manual issue 2.4 13 configuration menu parameters set zero parameter the set zero parameter is meant to be accessible to the user. It allows the removal of fixed display offset values from the display, so that the gross and net features can operate from a zero poi...

  • Page 14

    Interface 9320 teds ready indicator manual issue 2.4 14 auto off parameter the auto off parameter is another power saving feature. It allows for the setting of a time period in minutes, between 05 and 99 (00 de-activates auto off ). I.E. If this was set to 25, then if the 9320 detects no keyboard ac...

  • Page 15

    Interface 9320 teds ready indicator manual issue 2.4 15 calibration menu parameters sens 5.0 parameter the 9320 is factory set to enable calibration with sensors generating an input signal of 5mv/v or less. In the majority of cases it will not be necessary to read higher signal levels. If however, a...

  • Page 16

    Interface 9320 teds ready indicator manual issue 2.4 16 calibration procedures the best method of calibration, if it is possible to do so, is the live calibration, as this reads in the sensor signal at two calibration points and scales the 9320 automatically. If this is not possible, then the sensit...

  • Page 17: 5.0 Gain

    Interface 9320 teds ready indicator manual issue 2.4 17 internal calibration menu internal calibration of the 9320 is important when the indicator is used in the teds mode or when it is scaled by the table method. In either of these scenarios, the indicator either reads the transducer eeprom to get ...

  • Page 18

    Interface 9320 teds ready indicator manual issue 2.4 18 internal calibration procedure 1. Read and note the existing settings for gain and offset using the internal calibration menu. If desired, the settings for gain and offset may be reset to 1 and 0 respectively for convenience. 2. With the 9320 i...

  • Page 19

    Interface 9320 teds ready indicator manual issue 2.4 19 specifications performance input type: strain gage full bridge sensors input range: up ±5mv/v (±50mv/v can be supplied, with factory set option) nonlinearity: ±0.005% fs thermal drift: excitation voltage: 5vdc (±4%), 59ma maximum current minimu...

  • Page 20

    Interface 9320 teds ready indicator manual issue 2.4 20 warranty the 9320 is warranted against defective material and workmanship for a period of (1) one year from the date of dispatch. If the interface, inc. Product you purchase appears to have a defect in material or workmanship or fails during no...