NBN 9320 User Manual - Set
Interface 9320 TEDS Ready Indicator Manual Issue 2.4
9
This allows the setting of a visual overload. The value entered is the display value at which
the 9320 displays
OUErLOAd
.
Values between
-9999999
and
+9999999
can be entered, using the
and
arrows to
select a digit and the
and
arrows to increment or decrement the digits. Press
to accept the value and move onto the next parameter.
Parameter
Set-up Information
SEt
OPEr
Press
To skip to next menu item
Press
To select operation mode
This allows the enabling or disabling of the power save mode, which updates at 1 update per
second and pulses the sensor excitation. This results in a lower accuracy (1 part in 20,000).
To enable press
.
To disable press
AUtO
OFF
Press
To skip to next menu item
Press
To set auto power off
This enables the setting of an auto power off value. The value entered is in minutes. If no
front panel buttons are pressed for the time set here, then the indicator will automatically
power off, to conserve battery life.
Values between
05
and
99
can be entered (
00
leaves the 9320 permanently powered), using
the
and
arrows to select a digit and the
and
arrows to increment or
decrement the digits. Press
to accept the value and move onto the next parameter.
rS232
Press
To skip this parameter and exit menu
Press
To enable the RS232 output
This feature enables you to enable or disable the RS232 output. Further details of the RS232
format are provided further into this manual. The RS232 output is an option that has to be
ordered with the 9320. To conserve battery life, it is suggested that the RS232 output is
disabled, when it is not required.
To enable press
To disable press