NBN 9320 User Manual - Sens 5.0
Interface 9320 TEDS Ready Indicator Manual Issue 2.4
10
Calibration Menu
To enter the Calibration Menu, press and hold
and
buttons for 5 seconds
Parameter
Set-up Information
SEnS 5.0
Press
To skip to next menu item
Press
To change sensor input sensitivity
This allows the calibration engineer to change the sensitivity range of the 9320, when
connecting to sensors with a sensitivity of greater than 5mV/V. The 9320 is factory set for
5mV/V. To ensure the unit is set to 5mV/V press
To select 50mV/V you need to power down the unit and access the internal circuit board.
Move link LK1 and place it onto JP1. Power on the 9320 and return to this point of the
calibration menu. You will notice that the menu parameter has changed to
SEnS 50.0
, press
to change the sensitivity to 50mV/V and move on to the next parameter.
SEt rES
Press
to skip to next menu item
Press
to the set the display resolution
This parameter sets the decimal point position for the display and the resolution, i.e. a value
of
000.005
would display the reading to 3 decimal places and the readings will change in
steps of 0.005.
The decimal point position is moved one place to the right each time you press the
and
together.
Any value can be entered for the resolution, using the
and
arrows to select a digit
and the
and
arrows to increment or decrement the digits. Press
to accept the
value and move onto the next parameter.
To save the settings and move on to the next parameter press
CALibrAt
THIS MENU IS DISABLED WHEN TEDS IS ENABLED
Press
to skip to next menu item.
Press
to the enter the calibration routine
If you have chosen to enter the calibration routine you will be prompted whether you want
to select
LiVE
, if you do not press
,
otherwise press
. You will then be prompted to
select either of the other calibration methods, which in order, are
tAbLE
and
CAL VAL
to
select any of the calibration methods press
. Otherwise press
For more detailed calibration information, please refer to the calibration section of the
manual.