NBN 9320 User Manual - page 15
Interface 9320 TEDS Ready Indicator Manual Issue 2.4
15
Calibration Menu Parameters
SEnS 5.0
Parameter
The 9320 is factory set to enable calibration with sensors generating an input signal of 5mV/V or less. In the
majority of cases it will not be necessary to read higher signal levels. If however, a higher sensitivity sensor is used
with the 9320, it will be necessary to gain access to the internal PCB (you must turn the 9320 off) to move link LK1
to JP1 (see picture below) to allow the 9320 to accept sensitivities of up to 50mV/V. TEDS should only be used with
5mV/V as 50mV/V is not factory calibrated.
Once this link has been moved, you will need to go back into the CALIBRATION MENU. When re entering the menu,
you will notice that the parameter
SEnS 5.0
has changed to
SEnS 50.0
to change the sensitivity to 50mV/V press
, the 9320 will now check the position of the link and change the sensitivity. It will now be necessary to re-
calibrate any sensors that you may previously have calibrated to this instrument.
SEt rES
Parameter
This parameter enables the setting of two features on the 9320. It allows you to set the decimal point position of
the display, by pressing the
and
together, to move the point position (each press moves the decimal point
position, one place to the right).
It also allows for the setting of the display resolution or the number of display counts the display changes with an
input change. To change the resolution use the
and
arrows to select a digit you want to change and the
and
arrows to increment or decrement the digits. Press
to accept the value.
CALibrAt
Parameter (disabled when TEDS is enabled)
This parameter is used to calibrate and scale the 9320 with a sensor. There are two basic methods of calibration
available. These are
LiVE
and
tAbLE
. There is also a third parameter, which can be used for maintenance and
recording purposes. This parameter is
CAL VAL
. The
CAL VAL
value can be viewed after a calibration has been
completed and will show the offset and gain figures from any stored calibration. If these figures are noted, they can
be used to re-enter at a later date, if calibration data is lost for any reason, or if the calibration data from a sensor
needs to be transferred to another 9320.
tedS
Parameter
This parameter automatically calibrates the 9320 with the data from the TEDS chip. The two annunciators appear
when active connection with a TEDS peripheral has been made. When there is a loss of connection these
annunciators flash. When changing a sensor the 9320 should be power cycled as this is when the TEDS data is read.
Calibration Procedures are not available when TEDS is enabled.
Sensitivity link should be in
this position for use with
sensors, with sensitivities
Sensitivity link should be
in this position for use
with sensors, with
sensitivities >+/-5mV/V