NBN 9320 User Manual - Set Zero
Interface 9320 TEDS Ready Indicator Manual Issue 2.4
8
SEnS 5.0
SEt rES
LiVE?
0000.000
uSE SC?
APPLY LO
dISP LO
0000000
APPLY HI
dISP HI
0000000
donE
APPLY LO
dISP LO
0000000
dISP HI
0000000
donE
CALibrAt
tAbLE?
InPut LO
0000000
dISP LO
0000000
InPut HI
0000000
dISP HI
0000000
donE
CAL VAL?
SEt 9Ain
0000000
SEt OFFS
donE
0000000
RETURN TO
NORMAL DISPLAY
MODE
tedS
EnAbLEd?
*
* Note:Only when TEDS is disabled
Configuration Menu
To enter the CONFIGURATION MENU, press and hold
and
buttons for 3 seconds
Parameter
Set-up Information
SEt ZEro
Press
To skip to next menu item
Press
To set a new system zero
This allows the user to introduce a fixed offset to the display value. The GROSS and NET
values are then displayed with this offset taken into account.
Values between
-9999999
and
+9999999
can be entered, using the
and
arrows to
select a digit and the
and
arrows to increment or decrement the digits. Press
to accept the value and move onto the next parameter.
Set Zero may also be set by pressing and at the same time.
SEt rAtE
Press
To skip to next menu item
Press
To change the update rate
This allows the user to set the display update rate, the options available is the update rate
of the display in Hz. Please note that the 25Hz update is only available in PEAK or VALLEY
mode.
When you choose to change the update rate you will be prompted whether you want to
select 25Hz, if you do not press
you will then be prompted to select any of the other
values, which in order, are 10Hz, 3Hz, 1Hz, 0.5Hz. to set the update rate for the value you
want press
SEt
OUEr
Press
To skip to next menu item
Press
To set overload alarm