8. Use the “Up Arrow” button and the “Down Arrow” button to change the Actual
Torque to the value given by the transducer on the Calibration Stand.
9. Press the “Centre” button to deselect the Actual Torque.
10. Press the “Up Arrow” button to display the next Calibration screen.
11. Repeat steps 6 – 10 until a Torque Cycle has been conducted at 30% to 100%.
12. Press and hold the “Centre” button until the Calibration Menu screen is displayed
(Figure 3.3.12-5).
Figure 3.3.12-5: Calibration Menu Screen
13. Use the “Up Arrow” button and the “Down Arrow” button to highlight the “Save”
option.
14. Press the “Centre” button to select the “Save” option.
Note: To exit calibration without saving, select the “Exit” option.
3.3.13 Default Calibration
This function allows the operator to use the Default Calibration Values while calibrating the
tool (refer to Section 3.3.12 – Tool Calibration). The Default Calibration Values are a
reference point for efficient calibration.
Note: These values should never be used as Calibration Values. After enabling the Default Calibration
Values, the operator must calibrate the Digital B-RAD.
To enable the Default Calibration Values:
1. Press the “Centre” button.
Result: The Main Menu screen will be displayed (Figure 3.3.13-1).
2. Use the “Up Arrow” button and the “Down Arrow” button to highlight the “Setup”
option.
3. Press the “Centre” button to select the “Setup” option.
Result: The Setup Menu screen will be displayed (Figure 3.3.13-2).
Figure 3.3.13-1: Main Menu Screen
Figure 3.3.13-2: Setup Menu Screen