Rapiscan TSA TPM-903B Operation And Service Manual - page 20
TSA TPM903 Operations and Service Manual
Doc: # 5058 Rev. A
_____________________________________________________________________________________________________________________
_____________________________________________________________________________________________________________________
This document is the exclusive property of Rapiscan Systems.
Unauthorized use or distribution is strictly prohibited
14
If the HVDAC value reaches 1000 and target counts haven’t been achieved, the calibration will
fail because the HV is getting dangerously high (for the PMT). The unit will flash both LED’s,
emit a short
PILLAR 1
FAILED CALIBRATION
PRESS ANY KEY
After pressing any key, the user will be returned to the menu. (No sense in continuing with the
calibration at this point).
If the calibration completes successfully, the user will be prompted to place the source on pillar 2
and repeat the process for that pillar.
At the end of a successful calibration, the unit will flash both LED’s, emit a short
user will be presented with a message:
CALIBRATION
COMPLETED
PRESS ANY KEY
After user presses key:
PILLAR 1
HVDAC: X
PRESS ANY KEY
After user presses key the same info will be presented for pillar 2
Pressing any key at this point returns to menu.