Visual Inspection (First Test)
VISUAL INSPECTION HAS TO BE DONE BEFORE ANY OTHER TEST IS CARRIED OUT USING ANY OF THE RANGE OF TnP APPLIANCE
TESTERS
There is no damage or component defects to the accessories, plugs, outlet sockets or connectors (physical).
There are no cracks &/or abrasions.
There are no exposed inner cores or conductors (flexible) and the supply cords are not twisted or distorted.
Any Fuse / Over load protection components (if fitted) are checked.
All labels, markings and warning indicators (of the maximum load to be connected to the device) are legible and intact.
The insulation is not damaged in any way i.e. melted, cuts or abrasions. There are no iron filings in the insulation.
There is no insulation tape on the lead.
Any flexible cords and/or leads are effectively anchored (glands and grommets intact).
All covers or guards are in place and secure as intended by the supplier/manufacturer.
All safety devices and systems are in good working order. (i.e. overload latches & buttons).
No dust &/or dirt obstructs any exhausts or ventilation outlets.
All controls are working properly and are secure and aligned.
Important: If result is a FAIL!!
If any Equipment
FAILS
ANY
of the above, it should be deemed to have
FAILED
the Visual Test, and therefore no other tests
need be performed. If this is the case the Equipment should be tagged with a
DANGER TAG
and removed from service. It is
recommended by the manufacturer and distributor of this product that it
SHOULD NOT BE RETURNED TO SERVICE
. To do so
would be considered unsafe.
The Wavecom series of appliance testers have been designed & manufactured to exceed and comply
with the AS/NZ:3760 standard, and to aid the end user with simple everyday electrical symbols that
they can identify with which will allow them to start the testing regime quickly and efficiently, without
any complexity as illustrated below.