50 | www.eddyfi.com
Maintenance and Troubleshooting
System Recovery Method
1. Connect Reddy to a power outlet with the power cable.
2. Make sure that Reddy is off.
If it is not, turn it off.
3. Turn on the instrument.
4. Immediately and simultaneously press the probe nulling button and the change layout button (see
page 4) until the following appears.
Figure 6–1 Options menu
5. With the keypad arrows, select Enter Reddy System Recovery, and then press the Enter button
(see page 4).
You are prompted to wait until the following appears.
Figure 6–2 System recovery interface
6. Using the keypad arrows, select Install the factory EddyOS update, keep data.
7. When prompted, press the up arrow of the keypad.
The update process starts. This normally takes between 5 and 10 minute. When the process is
complete, the system restarts.
8. Activate Windows.
See Activating Windows on page 50 for details.
Activating Windows
Microsoft requires that you activate Windows to be able to use it. The activation process is automatic
when you connect Reddy to the Internet through an Ethernet cable. To proceed :
1. Make sure that Reddy is on and that Magnifi GO is running.
2. Connect an Ethernet cable to Reddy.
3. Connect the other end of your Ethernet cable to a network (local area network or other access
point).
Wait until a dialog box confirming the activation of Windows appears on screen. If you do not activate
Windows, every time you start Reddy, a message will remind you to do so. You have 30 days to activate
Windows before it locks up.