JDS Uniphase OLP-34 Operating Instructions Manual - 8 S
Manual is about: Optical Power Meter
of
86
28
OLP-34/-35/-38
JDSU
8 S
PECIFICATIONS
OLP-35
Adjustable wavelength
range
780 to 1650 nm,
in 1 nm steps
Calibrated
wavelengths
850 nm, 980 nm, 1300 nm,
1310 nm, 1490 nm,
1550 nm, 1625 nm
Photo diode
InGaAs
Fiber type
9/125 to 100/140
Power/Display range
-65 to +10 dBm
Resolution
0.01 dB, 0.001 µW
Max. permitted level
+16 dBm
Intrinsic uncertainty
1)
1) Under reference conditions: -20 dBm (CW), 1310 nm ±1 nm,
23 °C ±3K, 45 to 75% relative humidity, 9 to 50 µm test fiber,
ceramic end face
± 0.2 dB (± 5%)
Linearity
2)
2) -50 to +5 dBm from -5 to +45 °C.
0.06 dB
Overall measurement
uncertainty
3)
850 nm
1300, 1310 nm
1550 nm
1625 nm
3) From -5 to +45 °C.
± 0.35 dB ± 0.8 nW
± 0.30 dB ± 0.1 nW
± 0.25 dB ± 0.1 nW
± 0.35 dB ± 0.1 nW