A&D HV-200KGL-WP Instruction Manual - page 32
Page 30
HVG-WP Series
A
Function
table
Meaning and purpose
A
F67
The comparator is permanently activated and compares stable
values only, except those near zero.
A
A
Caution
The upper limit value (HI) must be greater than the lower limit value (LO).
The parameters of the upper limit value (HI) and the final value (HI) use the
same memory. The parameters of the lower limit value (LO) and the zero band
(LO) use the same memory.
The upper/lower comparator function, the simple batch function and the
full/dribble batch function can not be used at the same time because these
parameters use common memory.
11.1.Preparation (Setting Parameters)
Step 1 Turn off the display.
Press
the
ON/OFF
key while the
ZERO
key is pressed
and held. F1 is displayed.
Step 2 Press the
ENTER
key repeatedly until F6 is
displayed.
Step 3 Select a parameter of the comparator function
(
f6 0
~
f6 7
) with the
∧
key.
Step 4 Press the
ENTER
key to store the new parameter.
Step 5 Press the
F
key and the
ENTER
key to exit from the
function mode. The scale returns to the weighing
mode.