Abbott i-STAT 1 System Manual - Customizing The Handheld
Customizing the Handheld
The handheld has a variety of settings that can be customized to make data management and
compliance with laboratory regulations easier in large hospital settings. Customization options that
may be useful to the smaller hospital or physician’s office using the i-STAT System under a CLIA
Certificate of Waiver are included below.
To view the active Customization Profile:
1. Press to turn on handheld.
2. Press to change screen to Administration Menu.
3. Press (Customization).
4. Press (View).
5. Select the category to view (1 - 5).
6. Use or to scroll the options in each category.
Use to return to the Customization menu.
To change the active Customization Profile:
1. Press to turn on handheld.
2. Press to change screen to Administration Menu.
3. Press (Customization).
4. Press (Change).
5. Press (Enter password if applicable).
6. Select the category to change (1 - 7).
7. To change a setting, select the item by pressing the
number key corresponding to the item, then select
the setting. Use arrow keys to scroll through all items.
8. Turn handheld off to save new setting(s).
MENU
4
2
6
Art: 720660-00K, Rev. Date: 11-Oct-17
15:26 25FEBYY
Administration Menu
1 - Analyzer Status
2 - Data Review
3 - Quality Tests
4 - Customization
5 - Set Clock
6 - Transmit Data
7 - Utility
Customization
Default0
1 – View
2 – Change
15:26 25FEBYY
Administration Menu
1 - Analyzer Status
2 - Data Review
3 - Quality Tests
4 - Customization
5 - Set Clock
6 - Transmit Data
7 - Utility
Customization
Default0
1 – View
2 – Change
View
Customization
1 – Analyzer
2 – ID Entry
3 – Patient Tests
4 – QC Tests
5 – Results
Change
Customization
Default0
1 – Analyzer
2 – ID Entry
3 – Patient Tests
4 – QC Tests
5 – Results
6 – Password
7 – Restore Factory
Settings
MENU
4
1
ENT