Abbott i-STAT 1 System Manual - page 62
14
Art: 720660-00K, Rev. Date: 11-Oct-17
Printing with the i-STAT Printer
1. Ensure that the printer is turned on and that the POWER indicator is green.
2. Align the handheld’s IR communication window with the printer’s IR LED window. Generally, the
printer must be within 1 to 5 inches (2.5 – 12.7 cm) of and not too close to the handheld.
3. Display the results to be printed on the handheld.
4. Press on the handheld. Do not move handheld or printer until printing is complete.
5. If printer is not powered from a wall outlet using the AC adapter, turn printer off.
Printing Many Results
1. Press to turn on handheld.
2. Press to access the Administration Menu.
3. Press – Data Review.
4. Press – List.
5. Scroll through the results using the
← and → keys.
6. Press the number key for the test record(s) to be printed. (Press the numbered key again to
deselect a record.)
7. Align i-STAT 1 handheld and i-STAT Printer IR window. Press on the handheld.
8. Do not move handheld or printer until printing is complete.
9. If printer is not powered from a wall outlet using the AC adapter, turn printer off.
What is Printed:
Name of Test
i-STAT cartridge type.
Sample ID
Patient ID or type of quality test and lot number of solution.
Results
Results are printed with units as well as flags, reference ranges,
and comment codes if applicable.
Sample Type
Sample type selected from Chart Page when sample is patient or
proficiency test.
Free Fields
Information entered into the free fields on the Chart Page when
sample is patient or proficiency test.
Time and Date
Time and Date when test was performed.
Operator ID
Operator ID
Lot Number
Lot number of cartridge
Serial
Serial number of the handheld
Version
Handheld application software
CLEW
Standardization software
2
7
MENU