Abbott i-STAT 1 System Manual - General Quality Check Codes
Art: 720660-00K, Rev. Date: 11-Oct-17
33
Cause Message
Action Message
Code
Comments
Cartridge Type Not
Recognized
Use Another Cartridge
69
Check that you are using the most recent software. Some cartridges require the most recent software. Make sure cartridges being used are not expired.
During testing, this code will be displayed if incorrect information is entered in response to the prompt “Enter or Scan Cartridge Lot Number.” The Handheld expects the barcode on the individual cartridge pack to be scanned. It will not accept keypad entries of the cartridge lot number nor a scan of the barcode on the
cartridge box.
Cartridge Error
Use Another Cartridge
79-81
This may be a cartridge problem, but if the failure persists it is likely due to
damaged thermal probes in Handheld. Attempting to remove a cartridge or
simulator before the “Cartridge Locked” or “Simulator Locked” message is
removed or results are displayed will damage
thermal probes.
Test Cancelled by
Operator
95
This message will appear in the stored results if the Handheld powers
down before mandatory information was entered.
Analyzer Error
See System Manual
Various
Codes
Insert Electronic Simulator twice and test a control sample. If Handheld
passes simulator checks and control is in range, continue to use it. If not, call
Technical Support. Handheld may need to be replaced.
Electronic Simulator Fail
L
Allow Handheld to equilibrate in new environment for 30 minutes and
repeat test. If code reoccurs, call
Technical Support. Handheld may
need to be replaced.
Electronic Simulator Fail
G
Reinsert simulator. If code reoccurs, call Technical Support. Handheld may
need to be replaced.
Electronic Simulator Fail
R, r, t, B
Call Technical Support. Handheld may need to be replaced.