Power
There are three options for powering the i-STAT Printer:
• Using the AC adapter and power cord only,
• Using the Rechargeable Battery only, and
• Using the Rechargeable Battery with the AC adapter and power cord.
The i-STAT Printer can be turned on and off by pressing the POWER button. When the printer
is on, the POWER indicator will be illuminated:
Power OK:
Green
Battery Low:
Orange
Battery Empty:
Red
If the printer is inactive for >60 seconds, it will automatically enter the power-saving mode. When
in the power-saving mode, the POWER indicator will change from a solid color light to pulsed
illumination.
The printer’s rechargeable battery needs to be recharged when the POWER indicator turns
orange. If the battery becomes exhausted, the POWER indicator will turn red and printing will
be disabled.
The printer’s battery can be recharged using the supplied AC power adapter. The socket for the
AC power adapter is located on the rear of the printer. Note: Charging only occurs when the
printer is switched off or in the power-saving mode. A full charge takes approximately 3 hours.
Symptoms Indicating that the Rechargeable Battery Requires Replacement:
1. A steady Orange or Red POWER indicator light on the printer, even after charging it for the
recommended 3 hours.
2. Loss of battery capacity, indicated by a shorter interval between charges.
Installing or Replacing the Rechargeable Battery in the i-STAT Printer:
1. Disconnect the printer from the AC adapter.
2. Turn the i-STAT Printer upside down and place it on a flat surface. Remove the battery door
by sliding it off while pressing on the grooved section. Set the door aside.