Abbott i-STAT 1 System Manual - General Quality Check Codes
32
Art: 720660-00K, Rev. Date: 11-Oct-17
Cause Message
Action Message
Code
Comments
Dead Batteries
Replace Batteries
1
There is insufficient power to complete the testing cycle. Replace
the batteries.
Temperature Out of
Range
Check Status Page
2
Move the Handheld to an area within the operating temperature
range: 16-30 °C (61-86 °F). Check the Status page to see when the
temperature is within operating range.
Date Invalid
Check Clock on Status
Page
11
Date in Handheld precedes release date programmed into the
application software. Correct the date.
Expired Software
Update Required
See Manual
12
CLEW software has expired. Install most recent CLEW software. If
CLEW has not expired, check date on Handheld.
Invalid Software
Update Required
See Manual
13
CLEW software is corrupt or there is no CLEW software in
Handheld. Install or re-install most recent CLEW software.
Cartridge Not
Inserted Properly
Reinsert Cartridge
47
Push cartridge straight trough the cartri
dge door until it will go no
further.
Analyzer Error
Use Electronic Simulator
50-52,
58-62
If the Handheld passes the simulator check, continue to use it. If
not, contact Technical Support. Handheld may need to be replaced.
Analyzer error
See Manual
56
Interference from electronic noise from nearby power supply or
power cord. Relocate Handheld.