Accu-Chek Mobile User Manual - page 183
Other manuals for Mobile:
User Manual, Quick Reference Manual, User Manual, Getting Started Manual, User Manual, User Manual, User Manual, Getting Started Manual
Manual is about: BLOOD GLUCOSE MONITORING SYSTEM
of
255
183
Checking the meter 7
Values are only examples
or
The meter displays the upper and lower limits of the
concentration range. The concentration range dis-
played depends on the color of control solution se-
lected.
15
Press
(Continue)
.
The meter automatically checks whether the result is
within the concentration range displayed.
If the result is within the concentration range displayed,
Control test OK
is displayed.
If the result is not within the concentration range
displayed,
Control test not OK ... Repeat control test
will be displayed (see Chapter “Control test not oK –
possible sources of error”).
16
Press
(OK)
.