EDAN SONOTRAX Basic User Manual - page 28
SONOTRAX Series Ultrasonic Pocket Doppler User Manual Operation
- 23 -
3.9 Vascular Examining (Optional)
WARNING
The device is not intended for ophthalmic use. Do not use it for examining ophthalmic
vessels, or any other procedures which may cause the ultrasound beam to pass through
the eye.
4MHz, 5MHz or 8MHz vascular probes are to be connected to the Doppler to perform vascular
examination.
Choose the appropriate probe as required. The probe with how frequency has a deeper
penetration depth, while the probe with high frequency has better resolution and wider detecting
range. The 4MHz vascular probe is optimized for examining blood vessels; the 5MHz vascular
probe is optimized for examining deeper vessels, and the 8MHz vascular probe is optimized for
examining surface vessels.
Apply a liberal amount of gel on the site to be examined. Place the probe at a 45º angle on the skin
over the vessel to be examined. Adjust the position of the probe to obtain the loudest blood flow
sound. Refer to figure 3-13 for the probe sites: