EDAN SONOTRAX Basic User Manual - Manufacture’S Declaration
SONOTRAX Series Ultrasonic Pocket Doppler User Manual EMC Information
- 34 -
Appendix 2 EMC Information-Guidance and
Manufacture’s Declaration
A2.1 Electromagnetic Emissions - for all Equipment and
Systems
Guidance and manufacture’s declaration–electromagnetic emission
The SONOTRAX Series Ultrasonic Pocket Doppler is intended for use in the electromagnetic
environment specified below. The customer or the user of the device should assure that it is used in such
an environment.
Emission test
Compliance
Electromagnetic environment–guidance
RF emissions
CISPR 11
Group 1
The SONOTRAX Series Ultrasonic Pocket Doppler uses RF
energy only for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any interference in nearby
electronic equipment.
RF emission
CISPR 11
Class B
The SONOTRAX Series Ultrasonic Pocket Doppler is suitable for
use in all establishments, including domestic establishments and
those directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
A2.2 Electromagnetic Immunity - for all Equipment and Systems
Gu
idance and manufacture’s declaration–electromagnetic immunity
The SONOTRAX Series Ultrasonic Pocket Doppler is intended for use in the electromagnetic
environment specified below. The customer or the user of the device should assure that it is used in such
an environment.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic
environment-guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile. If
floor are covered with
synthetic material, the relative
humidity should be at least
30%.