Smartlink MEDI-CALL UNIT Installation Manual - 6.5 System Alarms
Medi-Call V2.0 – V2.7 Installation Manual
25
6.5 SYSTEM ALARMS
System alarms are indicated by flashing the alarm type on the status display. The speaker/beeper also beeps for 3 seconds.
NOTE: System alarms are not indicated when the system is ARMED or in HOME Mode.
Ademco and British DTMF Versions – Status display will keep flashing or stay On until the system alarm is corrected.
Contact ID Version - The OFF Button on Radio Key Pendant clears the display. Pressing the OFF Button again will
redisplay the alarm.
NOTE: The Display will turn Off 5 minutes after a Mains Fail is detected and will remain Off until Mains is Restored.
DISPLAY
DESCRIPTION
-----
LOW BATTERY RADIO KEY PENDANTS.
-----
LOW BATTERY RADIO DETECTORS.
-----
SUPERVISION FOR RADIO DETECTORS.
FAILED TO REPORT ON DIALLER.
LINE FAIL (Line Fail must be detected for 2 minutes before “LF” is
displayed and the SmartLink unit beeps for 25 seconds).
6.6 PRE-ALARM “PA”, ENTRY TIME “En” and EXIT TIME “Ei”
PRE-ALARM TIME “PA” allows a time between a Radio Key/Detector triggering and SmartLink alarming.
ENTRY TIME “En” allows a time from activating a detector till DISARM. During Entry Delay the siren will chirp once a
second. When the OFF Button is pressed the speaker beeps 3 times to indicate DISARMED. If the OFF Button is not
pressed within the Entry Delay, an alarm is generated.
EXIT TIME “Ei” allows a time from pressing the ON Button on a Radio Key Pendant until ARMED. The speaker will beep
once when the ON Button on a Radio Key Pendant is pressed to indicate Exit Delay start and beep once at exit delay end to
indicate ARMED.
6.7 ALARM LOCKOUT “Lc”
The SmartLink unit inhibits multiple activations of the same alarm. ie. the Open Collect Outputs (OUTPUTS “o1” – “o3”)
will not turn on again or the SmartLink unit will not report same alarm during the particular ARMED session when “Lc”
LOCKOUT programming option is Enabled.
NOTE: Lockout does not apply to PANIC, DURESS or DEMENTIA alarms or any Detectors with the “Fi” FIRE or “In”
INACTIVITY option. Lockout will also effect the Radio transmitted message as well.
6.8 INACTIVITY ALARM “In”
The SmartLink unit has an optional INACTIVITY ALARM (1-200=1-200 Hours and 201-255=1-55 Minutes). If the option
(“In” INACTIVITY TIMER) is Enabled, then the Inactivity Timer will be reset by the CANCEL (Green) Button. Inactivity
Detectors and Dementia Keys reset the Inactivity Timer if the timer has not expired. If the CANCEL (Green) Button is not
pressed or no movement is detected or no signals from Dementia Keys for the programmed INACTIVITY TIME, the
SmartLink unit will produce warning beeps and flash the HELP (Red) Button. PRE-ALARM TIME “PA” (0-60 Seconds)
before the programmed Inactivity Timer and an alarm is sent to the Control Centre. Depending on the program option “Ir”
INACTIVITY RESTORE, if Enabled the inactivity will restore when an activity message is received, otherwise it will not
restore until the CANCEL (Green) Button is pressed.
NOTE: If “SH” QUIET option is Enabled, then beeps are suppressed and if there is a call back, SmartLink unit answers
only in Listen Mode. Alarms are also Reported straight away to the Control Centre.
6.9 TIME TILL FIRST TEST “1C” & TEST CALL TIME ‘2C”
“1C” TIME TILL FIRST TEST CALL (0 – 24 Hours) should be programmed so that the first test call will be made during a
low activity period of the Control Centre. eg. midnight. By programming 0 then the SmartLink unit will report a test call
immediately after Exit Program Mode.
“2C” TEST CALL TIME can be programmed from 1 – 255 hours. By programming 0 then this will Disable the reporting of
the test call.