Vega VEGABAR 54 Product Information

Manual is about: pressure transmitter

Summary of VEGABAR 54

  • Page 1

    Pressure measurement process pressure/hydrostatic vegabar 51 vegabar 52 vegabar 53 vegabar 54 product information.

  • Page 2

    Contents 2 process pressure/hydrostatic – pressure measurement contents 1 application, function, configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 2 type overview . . . . . . . . . . . . . . . . . . . . . . . . ...

  • Page 3

    1 application, function, configuration area of application vegabar series 50 sensors are pressure transmitters with a deviation in characteristics of 0.2 % or 0.1 % for process pressure and level measurement. In their construction and level of accu- racy they are suitable for all standard application...

  • Page 4

    1 4 3 2 fig. 2: configuration of the piezoresistive measuring cell in vegabar 53 1 sensor element 2 base element 3 diaphragm 4 silicone oil filling the features of the piezoresistive measuring cell are: l elastomere-free l wetted parts of stainless steel l low hysteresis configuration vegabar series 50...

  • Page 5

    Chemical pump fig. 8: dry run protection on a chemical pump with vegabar 53 the vegabar 53 pressure transmitter with piezoresistive meas- uring cell and metal diaphragm is deployed as dry run protection system for chemical pumps. Its strengths are the small, front-flush process fitting, also for small...

  • Page 6

    2 type overview vegabar 51 vegabar 52 vegabar 52 - climate compensated measuring cell: certec ® /dms certec ® certec ® diaphragm: metal ceramic ceramic media: gases, vapours and liquids, also aggressive ones, at high temperatures gases, vapours and liquids gases, vapours and liquids process fitting: ...

  • Page 7

    Vegabar 53 vegabar 54 vegabar 54 - climate compensated measuring cell: piezoresistive/dms certec ® certec ® diaphragm: metal ceramic ceramic media: gas, vapours and liquids, also viscous gases, vapours and liquids, also with abrasive substances gases, vapours and liquids, also with abrasive substanc...

  • Page 8

    Indicating and adjustment module plicscom housing plastic stainless steel aluminium aluminium (double chamber) electronics 4 … 20 ma/hart profibus pa foundation field- bus process fitting thread flange sanitary sensors certec ® measur- ing cell piezoresistive measuring cell approvals sil overfill prote...

  • Page 9

    3 mounting instructions mounting position vegabar functions in any installation position. Depending on the measuring system, the installation position can influence the measurement. This can be compensated by a position correc- tion. Fig. 10: typical mounting arrangements for process pressure transmi...

  • Page 10

    4 electrical connection 4 .1 general requirements the supply voltage range can differ depending on the instrument version. You can find exact specifications in chapter "technical data". The national installation standards as well as the valid safety regulations and accident prevention rules must be obs...

  • Page 11

    4 .5 wiring plan single chamber housing i2c display 1 1 2 5 6 7 8 fig. 12: connection hart two-wire, profibus pa, foundation fieldbus 1 voltage supply and signal output double chamber housing - two-wire i2c 1 1 2 fig. 13: connection hart two-wire, profibus pa, foundation fieldbus 1 voltage supply and ...

  • Page 12

    5 operation 5 .1 overview the sensors can be adjusted with the following adjustment me- dia: l with indicating and adjustment module l an adjustment software according to fdt/dtm standard, e.G. Pactware and pc and, depending on the signal output, also with: l a hart handheld (4 … 20 ma/hart) l the a...

  • Page 13

    Ment family. This licence can be bought from the vega agency serving you. Connection of the pc via vegaconnect 3 1 2 fig. 16: connection of the pc via vegaconnect directly to the sensor 1 usb cable to the pc 2 vegaconnect 3 sensor 1 2 3 4 open twist usb lock fig. 17: connection via i²c connection ca...

  • Page 14

    6 technical data general data 316l corresponds to 1.4404 or 1.4435 common data materials, non-wetted parts - electronics housing plastic pbt (polyester), alu die-casting powder-coated, 316l - seal nbr (stainless steel housing), silicone (alu/plastic housing) - inspection window in housing cover for ...

  • Page 15

    Profibus pa output signal digital output signal, format according to ieee-754 - sensor address 126 (default setting) current value constantly 10 ma, ±1 ma integration time 0 … 999 s, adjustable foundation fieldbus output - signal digital output signal, foundation fieldbus protocol - physical layer ac...

  • Page 16

    Technical data 16 process pressure/hydrostatic – pressure measurement resolution 1 °c (1.8 °f) accuracy - in the range of 0 … + 100 °c (+32 … +212 °f) ± 3 k - in the range of -50 … 0 ° c (-58 … +32 °f) and +100 … +150 °c (+212 … +302 °f) typ. ±4 k input variable measured value process pressure measu...

  • Page 17

    Technical data process pressure/hydrostatic – pressure measurement 17 average temperature coefficient of the zero signal - turn down 1 : 1 0.05 %/10 k - turn down 1 : 1 up to 5 : 1 0.1 %/10 k - turn down up to 10 : 1 0.15 %/10 k outside the compensated temperature range: average temperature coefficient ...

  • Page 18

    Technical data 18 process pressure/hydrostatic – pressure measurement process conditions vegabar 51 product temperature depending on the isolating liquid (temperature: p abs > 1 bar/14.5 psi/p abs 1 bar/14.5 psi) 8) - silicone oil kn2.2 -40 … +150 °c/-40 … +150 °c (-40 … +302 °f/-40 … +302 °f) - sil...

  • Page 19

    Technical data process pressure/hydrostatic – pressure measurement 19 indicating and adjustment module voltage supply and data transmission through the sensor indication lc display in dot matrix adjustment elements 4 keys protection - unassembled ip 20 - mounted into the sensor without cover ip 40 m...

  • Page 20

    Technical data 20 process pressure/hydrostatic – pressure measurement 1000 750 500 250 14 18 16 20 22 24 26 28 30 32 34 36 Ω v 4 1 2 3 fig. 20: voltage diagram vegabar 53 1 hart load 2 voltage limit eex-ia instrument 3 voltage limit non-ex instrument 4 supply voltage voltage supply - profibus pa vega...

  • Page 21

    Technical data process pressure/hydrostatic – pressure measurement 21 supply voltage with lighted indicating and adjustment module - non-ex instrument 13 … 32 v dc - eex-ia instrument 13 … 24 v dc power supply by/max. Number of sensors - h1 fieldbus cable/voltage supply max. 32 (max. 10 with ex) ele...

  • Page 22

    Dimensions 22 process pressure/hydrostatic – pressure measurement 7 dimensions housing in protection ip 66/ip 67 ~ 69 mm (2 23 / 32 ") ø 77 mm (3 1 / 32 ") 112 mm (4 13 / 32 ") m20x1,5/ ½ npt ~ 69 mm (2 23 / 32 ") ø 77 mm (3 1 / 32 ") 117 mm (4 39 / 64 ") m20x1,5/ ½ npt ~ 87 mm (3 27 / 64 ") m16x1,5...

  • Page 23

    Dimensions process pressure/hydrostatic – pressure measurement 23 vegabar 51 - threaded version sw32 sw41 sw55 sw55 g¾b g1b 1½”npt g1½b ø 22 mm (0.87") ø 32 mm (1.26") 20 mm (0.79") ~105 mm (4.13") ~105 mm (4.72") ~105 mm (4.13") ~120 mm (4.72") ø 25 mm (0.98") ø 39 mm (1.54") 28 mm (1.1") ø 40 mm (...

  • Page 24

    Dimensions 24 process pressure/hydrostatic – pressure measurement vegabar 52 gj gn/gm gi/gt/gc gb gs gp/gl/gu/gk sw 27 mm (1 1 / 16 ") g½a g½a g¼a 23 mm ( 29 / 32 ") 7 3 mm (2 7 / 8 ") 7 0 mm (2 3 / 4 ") 1 7 mm ( 43 / 64 ") 8 5 mm (3 11 / 32 ") 1 7 mm ( 43 / 64 ") 3 mm ( 1 / 8 ") 2 0 mm ( 25 / 32 ")...

  • Page 25

    Dimensions process pressure/hydrostatic – pressure measurement 25 vegabar 54 - flange fitting 1 ea, fb, fm, fh, fi 3 15 / 32 " 4 1 / 64 " 5 7 / 16 " 1 / 8 " 1 / 8 " 1 / 8 " 4xø 45 / 64 " 4xø 45 / 64 " 8xø 45 / 64 " 4 21 / 64 " 4 59 / 64 " 6 19 / 64 " 5 29 / 32 " 6 1 / 2 " 7 7 / 8 " 45 / 64 " 25 / 32 "...

  • Page 26

    Dimensions 26 process pressure/hydrostatic – pressure measurement vegabar 54 - fitting for paper industry 2 vp vr ø 29,3 mm (1 5/32 ") 30 mm (1 3 / 16 ") 104,8 mm (4 1 / 8 ") g1" sw36 ø 24,7 mm ( 31/32 ") ø 29,3 mm (1 5/32 ") 30 mm (1 3 / 16 ") ø 24,7 mm ( 31/32 ") 104,8 mm (4 1 / 8 ") g1" sw36 fig. ...

  • Page 27

    Product code process pressure/hydrostatic – pressure measurement 27 8 product code vegabar 51 approval xx without xm ship approval cx atex ii 1g, 1/2g, 2g eex ia iic t6 cm atex ii 1g, 1/2g, 2g eex ia iic t6 + ship approval gx atex ii 1/2d,2d ip6x t process fitting / material fr flange dn25pn40 form ...

  • Page 28

    28 process pressure/hydrostatic – pressure measurement vegabar 53 approval xx without xm ship approval cx atex ii 1g, 1/2g, 2g eex ia iic t6 cm atex ii 1g, 1/2g, 2g eex ia iic t6 + ship approval gx atex ii 1/2d,2d ip6x t ux fm(ni)cl i,div2,gp abcd (dip)cl ii,iii,div1,gp efg uf fm(is)cl i,ii,iii, div...

  • Page 29

    Process pressure/hydrostatic – pressure measurement 29 29234 -en -081013.

  • Page 30

    30 process pressure/hydrostatic – pressure measurement 29234 -en -081013.

  • Page 31

    Process pressure/hydrostatic – pressure measurement 31 29234 -en -081013.

  • Page 32

    Vega grieshaber kg am hohenstein 113 77761 schiltach germany phone +49 7836 50-0 fax +49 7836 50-201 e-mail: info@de.Vega.Com www.Vega.Com you can find at www.Vega.Com downloads of the following l operating instructions manuals l menu schematics l software l certificates l approvals and much, much mor...