Accu-Chek Performa Connect User Manual - page 84
82
Meter Maintenance and Troubleshooting
8
• If the second test result does
not match how you feel,
perform a control test with
the control solution and a new
test strip.
• If the control result is within
the acceptable range, review
the proper testing procedure
and repeat the blood glucose
test with a new test strip.
• If the control result is not
within the acceptable range,
see the
Understanding
Out‑of‑Range Control
Results section in the
chapter
Control Tests.
A meter or test strip error has
occurred.
Your blood glucose may be
extremely high or a meter or a
test strip error has occurred.
• If your test result matches
how you feel, contact your
healthcare professional
immediately.
• If your test result does not
match how you feel, repeat
the blood glucose test. See
the
Unusual Blood Glucose
Results section in the chapter
Blood Glucose Tests.
• If the E‑3 code still appears
for your blood glucose test,
your blood glucose result
may be extremely high
and above the system’s
reading range. Contact your
healthcare professional
immediately.
52135_07135203001_Inside_EN.indb 82
4/22/14 7:32 AM