LabStyle Dario User Manual - Housing
Chapter 4 - Testing Your Blood Glucose Level
step Five - returning the Meter to the dario™ housing
40
14. Return the meter to the housing as described in the next section.
sTeP Five - reTurninG The MeTer To The dario™
housinG
To return the meter to the dario™ housing:
1. Disconnect the meter from your smart mobile device.
2. Hold the meter at a angle with the part that connects to the smart mobile device
facing away from the ridged orange panel.
3. Push the meter into the housing at an angle and then push downwards until the
meter clicks into place.
TesT LiMiTaTions
Normal endogenous (within body) natural levels of uric acid, ascorbate (vitamin
C), bilirubin, triglycerides, and hemoglobin do not interfere with your blood glucose
results. Elevated levels of these compounds in your blood may, however, interfere
with the accuracy of your blood glucose results obtained with the Dario™ System.
The system should not be used if blood levels of these compounds exceed the
limitations detailed in the table below.
Compound
Limitation
Ascorbic acid
> 4 mg/dL
Uric acid
> 9 mg/ dL
Acetaminophen
> 6 mg/dL
Total bilirubin
> 40 mg/dL
Triglycerides
> 1,026 mg/ dL
The drugs L-dopa, dopamine, methyl-dopa, acetaminophen, and ibuprofen will
not interfere with Dario™ system blood glucose results when taken at therapeutic
concentrations.
Patients undergoing oxygen therapy may yield false results.