LabStyle Dario User Manual - Specifications
How the Glucose Meter Works
Chapter 7 - Technical Information & Specifcations
48
CHapTer 7 – TeCHnICal InforMaTIon &
SpeCIfICaTIonS
This section provides technical information on your Dario™ System.
HoW THe GluCoSe MeTer WorkS
The Dario™ system determines the amount of glucose in fresh capillary blood
using electrochemical biosensor technology. An electrical current is generated by
the reaction of glucose with the reagent on the test strip (glucose oxidase). The
strength of the current produced by the reaction depends on the amount of glucose
in the blood sample. The glucose meter detects the current and converts it into a
blood glucose reading.
CoMpaTIbIlITy
Dario™ is compatible for use on smart mobile devices. For full list of compatible
devices please refer to www.mydario.com/support/getting-started/
TeCHnICal SpeCIfICaTIonS
Meter Specifcations
Result Range
10-600 mg/dL (0.6-33.3 mmol/L)
Calibration
Plasma-equivalent
Sample
Fresh capillary whole blood
Test Sample
0.3 microliter
Test Time
6 seconds
Assay Method
Glucose Oxidase Biosensor
Hematocrit Range
20% - 60%