Magnacoustics Genesis ULTRA Installation And Operation Manual - page 12
Magnacoustics Inc™ © 3/2010 rev – j 12 |
Genesis ULTRA Installation and Connections
- continued
For Excite or newer systems go to step 13, For MR750 go to step 14
12: For Horizon LX locate
Run 788 and remove the DB25 from J7 on the WIM Module and
connect to the DB25 on the AutoVoice Breakout Box, then connect “CB4” to the other
DB25 on AutoVoice Breakout Box, and connect the other end of “CB4” to
J7 on the WIM
Module.
13: For Excite or newer systems locate
Run 1085 labeled OW1-A21-J7 and remove the DB25
from
J7 on the Wim Module and connect to the DB25 on the AutoVoice Breakout Box, then
connect “CB4” to the other DB25 on AutoVoice Breakout Box, and connect the other end of
“CB4” to
J7 on the WIM Module.
14: For MR750 systems locate
Run 3031 and remove the DB25 from J7 on the Audio Board in
the GOC and connect to the DB25 on the AutoVoice Breakout Box, then connect one side
of cable “CB4” to the other DB25 on AutoVoice Breakout Box, and connect the other end of
“CB4” to
J7 on the Audio Board.
15: Connect the female side of cable “CB8” 6ft AC Adapter Cable to
J12 on the rear of
Genesis Ultra module. Ensure that the power entry receptacle is set up for proper voltage
either 110 or 220 default is 110.
CAUTION: Please ensure that all connections are
complete before connecting AC power cord.
16: Connect FM and AM Antenna to the Stereo. Reference Antenna Connections in the
Stereo Instructions for proper installation of FM / AM Antenna.
17: For GE MRI’s remove block off plate from J53 or J52 at penetration panel. All others use
supplied kit or fabricate a DB9 slot for filter as required.
18: Install RF filter with
female toward computer room.
19: Locate existing electrical conduit between control console and computer room.
CAUTION:Assistance may be necessary to ensure Trauma free installation of cable CB1
20: Route and install Cable “CB1” (90 ft. DB9 Male to Female) with Female connector toward
Genesis Ultra module and Male toward penetration panel.
21: Connect “CB1” Male to the RF filter.
22: Zip tie cable “CB1” to existing supports, or cables from penetration panel to provide strain
relief.
23: Connect Female end of cable “CB1” to
J11 on the rear of Genesis Ultra module.
24: From the scan room side of penetration panel, connect cable “CB2” (50ft. DB9 Female to
Female) to the RF filter you installed on penetration panel.
25: Zip tie cable “CB2” to existing cables from penetration panel to provide strain relief.
26: Route cable “CB2” on opposite side of the cold head (if used) to the front of the MRI
utilizing existing conduits.