Magnacoustics Inc™ © 3/2010 rev – j 18 |
Directions for installing or re-installing an older
Magnacoustics system on a newer GE MRI continued
1.
Depress the 4 gray buttons around the upper perimeter of the Genesis Ultra module,
releasing the 4 pop-up tabs on the top of the chassis that conceal the securing screws
.
2.
Remove the 4 larger screws with the #2 Philips and set aside.
3.
The module separates like a clam shell, use an even force pulling up on the top section
alternating between the left and right side repeatedly until the top is free and set aside.
(Please Note; Be Sure the front face stays with the bottom half of the chassis)
4.
Look towards the back right of the chassis, use this pictorial to locate the cable going to J5
remove the molex connector and using pin-poker depress tab to release pins, upgrade /
reconfigure to this spec.
1
3 O'CLOCK
6 O'CLOCK
9 O'CLOCK
J6
^
1
2
3
4
Con
2
Back of
Chassis
4
J5
^
1
2
3
4
ACV
IN
12 O'CLOCK
MD
18
- DS
P - Re
v -
3
>
Amp
Located on
Back of
Chassis
Front of
Chassis
5.
Reassemble by reversing steps 3 and 2 then depress the popup tabs on the top of the
chassis.
6.
Reconnect all wires and confirm that AVC appears in the lower left hand corner of display on
our Tech. Unit when the GE console Mic button is depressed.
7.
If you require any further assistance please contact Magnacoustics Technical Support @
800-637-2282 (9am to 5pm M-F EST).