Directions for installing or re-installing an older
Magnacoustics system on a newer GE MRI
If the serial number of your Magnacoustics system is below 18904 and you are installing or re-
installing this system on an Excite, HD, HDE, HDX, HDXT, MR750 or newer MRI, a cable pin-
out upgrade might be required. To determine if the upgrade is necessary, complete this install
according to the previous instructions. Depress the GE console
Mic button while observing the
lower left hand corner of the display on our Technologist Control Unit; “AVC” should appear. If
“AVC” does not appear, then the upgrade will be required. (Please Note: Magnacoustics Serial
#18904 shipped in July of 2006)
Mean Time to Upgrade;
Approximately 15 minutes
Tools required;
#2 Philips
Small pin-poker or small screwdriver
Please follow the Important Safety Instructions on page 2 and confirm that
the power cord is removed from the power entry module before proceeding
to Step 1