5 Set up with the display and adjustment module
VEGABAR 81 • Slave sensor for electronic differential pressure
48050-EN-160212
2
1
100%
0%
5 m
(196.9
")
4
3
Fig. 8: Parameter adjustment example "Min./max. adjustment, level measure-
ment"
1 Min. level = 0 % corresponds to 0.0 mbar
2 Max. level = 100 % corresponds to 490.5 mbar
3 VEGABAR 81
4 VEGABAR 81 - Slave sensor
If these values are not known, an adjustment with filling levels of e.g.
10 % and 90 % is also possible. By means of these settings, the real
filling height is then calculated.
The real product level during the adjustment is not important, because
the min./max. adjustment is always carried out without changing the
product level. These settings can be made ahead of time without the
instrument having to be installed.
1. Set the requested damping of the output signal in the menu item
"
Damping
".
2. Select the output characteristics in the menu item "
Current out-
put
".
To detect parameterization errors reliably, safety-relevant parameters
must be verified before saving them into the instrument.
1. Enter PIN
In delivery status, the PIN is "0000".
2. Character string comparison
You then have to carry out the character string comparison. This is
used to check the character presentation.
3. Serial number acknowledgement
Afterwards you confirm that the serial number of your instrument was
carried over correctly. This is used to check device communication.
4. Verify parameters
Confirm the modified values one after the other.
Further steps
Lock setup - adjustment