Summary of LM1
Page 1
User manual lm1 3 clean-trace ™ luminometer.
Page 2: Clean-Trace
Ii (english) en 3 issue date: 2016-06 clean-trace ™ luminometer contents intended use......................................................................... 1 product description ..................................................... 3 using atp detection for hygiene monitoring and management ........
Page 3
1 (english) en intended use the 3m clean-trace luminometer is a battery operated luminometer utilized primarily for objective and sensitive determination of the end result of adenosine tri-phosphate (atp) tests. The 3m clean-trace luminometer power supply is provided to charge the battery. The 3m™ c...
Page 4
2 (english) en environmental operating conditions environmental condition operating condition units indoor use only - only connect the data connections to a computer that conforms to iec/en/ul/csa 60950-1 - keep 3m clean-trace luminometer away from liquids and high humidity when connected to the com...
Page 5
3 (english) en limitation of liability unless prohibited by law, 3m will not be liable to user or others for any other direct, indirect, consequential, incidental or special damages including, for example, lost profits, business, investments, or opportunities even if 3m has been advised of the possi...
Page 6
4 (english) en operating the 3m clean‑trace luminometer battery usage and maintenance • a 3m clean-trace luminometer has an internal rechargeable lithium ion battery that must be charged to use a 3m clean-trace luminometer. Charging the battery • select the charger adapter that is appropriate for po...
Page 7
5 (english) en switching the power off 1. Press the on/off button for 3 seconds. 2. [power off] appears to confirm the power off process has started. 3. Tap [power off]. 4. A confirmation message your luminometer will shut down appears. 5. Tap [ok]. The display screen goes black to confirm the power...
Page 8
6 (english) en figure 3 shows the step-by-step procedure for inserting a 3m clean-trace test into a 3m clean-trace luminometer for testing, and removing it after testing is completed. 1. Open the 3m clean-trace luminometer by pressing the rocker cap. 2. Insert an activated 3m clean-trace test into t...
Page 9
7 (english) en logging in on the 3m clean‑trace luminometer with a pin when the 3m clean-trace luminometer is turned on, it automatically does a diagnostic check. If the diagnostic check is within the required limits, the login screen appears. 1. The default user is automatically logged in. Tap [log...
Page 10
8 (english) en performing tests on test points within sample plans performing an atp test on a test point within a sample plan 1. To start a new run, tap the sample plan, then tap [new test run…]. • if a sample plan does not appear, tap scroll up or scroll down or swipe the list up or down to displa...
Page 11
9 (english) en 5. To remove the test from the 3m clean-trace luminometer, press the rocker cap, grasp the top of the test and pull upwards. Follow the instructions for disposal in the 3m clean-trace test instructions for use. Performing a retest on an atp test point with a caution or a fail test sta...
Page 12
10 (english) en • to add a comment to the result, tap add comment . The add comments screen and any previously added comments appears. Tap [add comments] then tap comment to add. Tap [ok]. Tap [back]. If you decide not to add a comment, tap [back] to return to the previous screen. 3. To test another...
Page 13
11 (english) en changing temperature or ph recording for a test point within a sample plan 1. To select the test point you want to retest if you are still in the same sample plan, tap next or previous . 2. To use a different sample plan, tap home , then tap the sample plan, then tap the date and tim...
Page 14
12 (english) en setting up a custom test type on a test point within a sample plan if a sample plan does not appear, tap scroll up or scroll down or swipe the list up or down to display additional sample plans. • if you are restarting a sample plan that you did not complete, tap the sample plan with...
Page 15
13 (english) en performing testing on unplanned test points performing an unplanned atp test 1. Tap [unplanned test] . To start a new run, tap [new test run…]. • or, if you are restarting a run of unplanned tests that you did not complete, tap the starting date and time you want to restart. 2. Tap s...
Page 16
14 (english) en synchronizing the 3m clean‑trace luminometer with the computer database use the sync manager software to initiate a synchronization between the 3m clean-trace luminometer and your computer. Refer to the using the sync manager section of the 3m clean-trace hygiene management software ...
Page 17
15 (english) en 4. Turn wifi on, and select your wifi network. If it’s a secure wifi, enter your security credential. Obtain this information from your it support. 5. Tap return . 6. Tap settings . 7. Tap [preferred connection - (connection type)] 8. Tap [wifi] 9. Tap [web service setup]. 10. Confir...
Page 18
16 (english) en changing the connection type 1. Tap settings . 2. Tap [preferred connection - (connection type)] 3. Tap the connection type you prefer: • usb • wifi • bluetooth • skip sync – this selection disables the sync process 4. Tap [select]. Setting synchronization time 1. Tap settings . 2. T...
Page 19
17 (english) en displaying software versions 1. Tap settings . 2. Tap [about]. 3. Swipe the list of software versions up to display the complete list. 4. Tap [continue]. Troubleshooting this section is intended to help you solve problems that you might encounter when using the 3m clean-trace luminom...
Page 20
18 (english) en synchronization failure instrument not registered to register an instrument the first time you synchronize it, log in to the sync manager on your computer, then synchronize the instrument. Refer to the using the sync manager section of the 3m clean-trace hygiene management software u...
Page 21
19 (english) en appendix the 3m clean-trace luminometer display screen is laminated, bonded and shatter-resistant. It is designed to hold together in the event of breakage. The 3m clean-trace luminometer allows for gloved operation. Gloves need to be 4 mil thick or less and cannot be woven with meta...
Page 22
3m health care 2510 conway ave st. Paul, mn 55144 usa www.3m.Com/foodsafety © 2016, 3m. All rights reserved. 3m and clean-trace are trademarks of 3m. Used under license in canada. All other trademarks are the property of their respective companies. 34-8718-2963-5 3 3m food safety 3m united states 3m...