3M LM1 User Manual - page 15
13
(English)
EN
PERFORMING TESTING ON UNPLANNED TEST POINTS
Performing an Unplanned ATP Test
1. Tap
[Unplanned Test] .
To start a new run, tap
[New Test Run…].
• Or, if you are restarting a run of unplanned tests that you did not complete, tap the starting date and time you
want to restart.
2. Tap Select Test Type
then tap the test type:
• UXL100(Surface ATP)
• AQF100
• AQT200(water total)
3. Tap
[OK].
4. Follow the Instructions for Use for the 3M Clean-Trace Test to collect a sample, activate the test and shake it for at least
5 seconds.
5. Open the 3M Clean-Trace Luminometer by pressing the rocker cap. Then insert the activated test into the chamber.
Keep the 3M Clean-Trace Luminometer upright whenever a test is in the chamber, to ensure accurate results and avoid
spillage of reagents, which could damage the sensitive components of the instrument.
6. Tap
[Start] to begin measuring. A progress bar
appears and moves from left to right. The test result appears
and is stored in the memory so it can be synchronized with the 3M Clean-Trace Hygiene Management Software.
7. To remove the test from the 3M Clean-Trace Luminometer, press the rocker cap, grasp the top of the test and pull
upwards. Follow the instructions for disposal in the 3M Clean-Trace Test Instructions for Use.
8. To do another unplanned ATP test, tap
[New] and repeat this process.
• Or, to do a retest, tap
[Retest]
then tap
[Start]
.
Performing an Unplanned Visual Inspection Test
1. Tap
[Unplanned Test] .
2. To start a new run, tap
[New Test Run…].
Or, if you are restarting a run of unplanned tests that you did not complete, tap the starting date and time you want
to restart.
• If a starting date and time does not appear, tap scroll up
or scroll down
or swipe the list up or down to
display additional Sample Plans.
3. Tap Select Test Type
then tap
[Visual Inspection].
4. Tap
[OK].
5. Tap
[PASS] for a Pass result or tap [FAIL] for a Fail result.
6. If you selected Pass, the Pass
display appears. If you entered Fail, the Fail display appears, and the add comment
symbol appears and
[Retest]
appears.
7. To do another Visual Inspection test, tap
[New].
• To add a comment, tap Add Comment
. The Add Comments screen and any previously added comments
appear. Tap
[Add Comments], then tap the comment to add. Tap [OK]. Tap [Back].
• If you decide not to add a comment, tap
[Back] to return to the previous screen.