3M LM1 User Manual - page 12
10
(English)
EN
• To add a comment to the result, tap Add Comment
. The Add Comments screen and any previously added
comments appears. Tap
[Add Comments] then tap comment to add. Tap [OK]. Tap [Back]. If you decide not to add a
comment, tap
[Back] to return to the previous screen.
3. To test another Test Point, tap Next Test Point or Previous Test Point until the Test Point you want appears.
• Or, to do a retest, tap
[Retest]
then tap
[Start]
.
• Or, to display the list of Test Points, tap List View
.
Recording Temperature or pH for a Test Point within a Sample Plan
1. To start a new run, tap the Sample Plan, then tap
[New Test Run…].
• If a Sample Plan does not appear, tap scroll up
or scroll down
or swipe the list up or down to display
additional Sample Plans.
• If you are restarting a Sample Plan that you did not complete, tap the Sample Plan with the starting date and time
you want. If that Sample Plan does not appear, tap scroll up
or scroll down
to display additional Sample
Plan names.
• The List View appears, showing a list of the Test Points in the Sample Plan, and the number of Test Points in each
result category:
Total number of
Test Points
Number in
Pass
category
Number in
Caution
category
Number in
Fail
category
• To select a Test Type, tap
[Test types…] then tap the Test Type to be performed. Tap [Done].
2. To select a Test Point to be tested, tap the Test Point.
• If a Test Point includes an ATP test and a Temperature or pH recording, the Temperature or pH recording appears
immediately after the ATP test.
• You can test points in the Sample Plan in sequence or in any order. The first Test Point in the Sample Plan appears on
the screen. If you want to test a different Test Point, tap Next Test Point until the test point you want appears. You
can display a previous test point by tapping Previous Test Point .
• To display an image and a description of the Test Point, tap information
next to the Test Point name. Tap
[Back]
to return to the previous screen.
3. Tap the field. Use the number pad to enter the Temperature test result or pH test result.
4. The test result appears and is stored in the memory so it can be synchronized with the 3M Clean-Trace Hygiene
Management Software. The result is also interpreted using your established pass/fail limits and is displayed on the
screen as PASS
, or FAIL .
• To add a comment to the result, tap the Add Comment
symbol . The Add Comments screen and any
previously-selected comments will appear. Tap
[Add Comments], and then tap the comment to add. Tap [OK].
Tap
[Back].
• If you decide not to add a comment, tap Back to go to the previous screen.
5. To test another Test Point, tap Next Test Point or Previous Test Point until the Test Point you want appears.
• Or, to do a retest, tap
[Retest]
then tap
[Start]
.
• Or, to display the list of Test Points, tap List
View
.
Adding Optional Comments to a Caution or Fail ATP Test Result
1. To view or add comments for a displayed test result, tap Add Comment
. The Add Comments screen and any
previously-selected comments appear.
2. Tap
[Add Comments], and then tap the comment to add. Tap [OK]. Tap [Back].
3. If you decide not to add a comment, tap
[Back] to go to the previous screen.