The 3M Clean-Trace Luminometer display screen is laminated, bonded and shatter-resistant. It is designed to hold together
in the event of breakage.
The 3M Clean-Trace Luminometer allows for gloved operation. Gloves need to be 4 mil thick or less and cannot be woven
with metallic substances.
Cleaning and Decontaminating the 3M Clean‑Trace Luminometer
WARNING: To reduce the risks associated with hazardous voltage, always turn the instrument off and disconnect from
outlet before cleaning the instrument.
Clean and decontaminate the instrument immediately if any spills occur.
Clean the external surfaces of the instrument regularly once per month.
Cleaning and Decontaminating the External surfaces:
1. Turn off the power by pressing and holding the on/off button for 3 seconds (1, see figure below), and then disconnect
the USB cable (2, see figure below) on the side of the instrument.
2. Use a disposable towel lightly dampened with a solution of mild, non-staining disinfectant/cleaner such as mild
dishwashing soap to gently wipe the outer surfaces of the instrument. Wring out the disposable towel thoroughly so
it is damp, but not dripping. Avoid the USB cable connection on the side of the instrument when cleaning the exterior
surfaces.
3. Use another disposable towel to repeat the procedure with DI water (distilled water or deionized water).
4. Allow the exterior surfaces of the instrument to air dry for at least 1 hour. Do not reconnect the power cable and the
USB cable until you have allowed the instrument to air dry for at least 1 hour.
5. Connect the USB cable (2), and turn on the power by pressing and holding the on/off button for 3 seconds (1).