3M LM1 User Manual - page 16
14
(English)
EN
SYNCHRONIZING THE 3M CLEAN‑TRACE LUMINOMETER WITH THE COMPUTER DATABASE
Use the Sync Manager software to initiate a synchronization between the 3M Clean-Trace Luminometer and your
computer. Refer to the Using the Sync Manager section of the 3M Clean-Trace Hygiene Management Software User
Manual for more information.
• You must synchronize the first time using the USB connection.
• A Diagnostic Check is automatically done when a synchronization is initiated.
• If you start a synchronization less than two minutes after completing the last previous synchronization, a warning
message appears. Wait at least two minutes, then start a synchronization again.
• If you synchronize a 3M Clean-Trace Luminometer that already contains Sample Plans that are different than the Sample
Plans on the computer you are currently synchronizing with, a warning message appears.
o Tap Settings. Tap
[Synchronize]. The progress bar shows when the synchronization is complete.
o To cancel the synchronization, tap
[Skip Sync]
o To continue the synchronization, tap
[Continue]. This option removes the Sample Plans and test results from the 3M
Clean-Trace Luminometer and replaces the Sample Plans with the Sample Plans on the computer with which you are
currently synchronizing.
• Synchronizing the 3M Clean-Trace Luminometer with the computer sends information in both directions:
Sent from Computer to 3M Clean-Trace Luminometer:
- User names and PINs
- Sample Plans and Test Points including pass/fail limits and images of the Test Points if available
- Organization configurations such as number of days to retain synchronized test results
Sent from 3M Clean-Trace Luminometer to Computer:
- Test Results including date and time of test, User who did the test and comments if available
- The 3M Clean-Trace Luminometer provides three modes of communication: USB cable, Wi-Fi connection and
Bluetooth connection
Setting Up Synchronization with a USB Connection
To use a USB cable connection: connect one end of the 3M provided USB cable to the USB/Power port and connect the
other end of the USB cable to a USB port on your computer.
• To register a 3M Clean-Trace Luminometer the first time you synchronize it, log in to the Sync Manager. Refer to the
Logging in Your User Name on the Sync Manager section of the 3M Clean-Trace Hygiene Management Software User
Manual.
• 1. Tap
[Settings]
. Tap
[Preferred Connection – (connection type)]. Tap [USB]. Tap [Select].
Setting Up Synchronization with a Bluetooth Connection
• You must synchronize the first time using the USB connection to register the 3M Clean-Trace Luminometer.
• To use a Bluetooth connection: set up the Bluetooth connection for the 3M Clean-Trace Luminometer. This procedure
requires simultaneous use of both a 3M Clean-Trace Luminometer and the Sync Manager software, which is installed
on your computer by the software installation wizard for the 3M Clean-Trace Hygiene Management Software. For
instructions on pairing a Bluetooth connection, refer to the Setting Up a Bluetooth Connection section of the 3M
Clean-Trace Hygiene Management Software User Manual.
Setting Up Synchronization with a WiFi Connection
• You must synchronize the first time using the USB connection to register the 3M Clean-Trace Luminometer.
• To use a Wi-Fi connection: set up Web Service on the 3M Clean-Trace Luminometer, which then sends the web service
information you set up to the computer the first time you synchronize using the Sync Manager software.
1. Touch the top of the screen for 1 second, then swipe down.
2. Tap Connections
.
3. Tap WiFi
.